抽象的な

Cleaning method development and validation of Brimonidine Tartrate Residue from Brimonidine Tartrate API

Sayyad. Kousr Ali*, Tanmoy Mandal


 

 

To Develop the Precise, Linear and Accurate Cleaning method Validation for the Brimonidine Tartrate from Brimonide tartrate Active Pharmaceutical Ingredient as per International Conference on Harmonization guidelines by using High Performance Liquid Chromatography.

Material and Methods:  In this method uses a reverse phase C18 column make as inert sustain AQ C18 column (250*4.6mm,5µ), mobile phase-a as 0.1% ortho phosphoric acid and mobile phase-b as acetonitrile : methanol (50:50) v/v , flow rate as 1.2 ml/min, and detection of wavelength AS 248 nm, column temperature 40°C, injection volume 10µL and autosampler temperature 25°C by using Photo Diode Array Detector.

Results: In this method Brimonidine tartrate peak was eluted at 3.9 minutes. Brimonidine tartrate linear range was 0.5 µg/mL to7.5 µg/mL. In System suitability the % Relative standard deviation observed as 0.5. In swab and Rinse samples recovery was found between  97 to 103%. The limit of Quantification and limit of Detection range is found to be 0.5µg/mL and 0.15 µg/mL respectively.

Conclusion: An accurate, linear and precise Reverse phase High Performance Liquid Chromatography method was developed and validated as per International Conference on Harmonization guidelines. Hence this cleaning method validation is used for routine analysis of the residue samples in various pharmaceutical manufacturing areas. 


免責事項: この要約は人工知能ツールを使用して翻訳されており、まだレビューまたは確認されていません

インデックス付き

  • キャス
  • Google スカラー
  • Jゲートを開く
  • 中国国家知識基盤 (CNKI)
  • サイテファクター
  • コスモスIF
  • 電子ジャーナルライブラリ
  • 研究ジャーナル索引作成ディレクトリ (DRJI)
  • 秘密検索エンジン研究所
  • ICMJE

もっと見る

ジャーナルISSN

ジャーナル h-インデックス

Flyer

オープンアクセスジャーナル