抽象的な

Mediated microbial biosensor using Bacillus subtilis for wastewater BOD measurement

Li Jie, Yang Ruibing, Chen Kun, Song Zhiwen


In order to get some efficient microbial membrane, a Bacillus subtilis (BA1304) was isolated from wastewater, and was identified by the MicroStation System, and was used as the sensitive material for the BOD sensor (BODs) after being immobilized. Three immobilized mediators: polyethylene, sodiumalginate and polyacrylamidewere tested for choosing proper cell carriers. Furthermore, four temperature grads (25ï‚°C, 30ï‚°C, 35ï‚°C and 37ï‚°C) were studied to find the optimal working conditions. The result showed that the polyethylene was proven to be the most efficientmediator to the electrode in the presence of excess glucose/glutamic acid (GGA). The effective culture temperature was 35ï‚°C for Bacillus subtilis by testing values in standard BOD solutions, which were estimated using the BODs with deviation less than 5%. The BOD values were determined using the sensor and compared favorablywith those determined by the conventional BOD5method.


インデックス付き

  • キャス
  • Google スカラー
  • Jゲートを開く
  • 中国国家知識基盤 (CNKI)
  • サイテファクター
  • コスモスIF
  • 電子ジャーナルライブラリ
  • 研究ジャーナル索引作成ディレクトリ (DRJI)
  • 秘密検索エンジン研究所
  • ICMJE

もっと見る

ジャーナルISSN

ジャーナル h-インデックス

Flyer

オープンアクセスジャーナル