抽象的な

RP HPLC determination of benzhexol hydrochloride in tablet formulations and urine

Rabie S.Farag, Ashraf M.Ahmed


Anew, rapid and sensitive reverse phase HPLC method was developed and validated for the determination of benzhexol hydrochloride in tablet formulations and urine. The mobile phase used acetonitrile and water, (50 %v/v) adjust pH to 5 using phosphoric acid. The separation was achieved on C18 reversed-phase column (250 mm4 mmi.d.). The flowrate was 0.6 ml/min and UV detection at 254nm. The retention time for benzhexol hydrochloride was 7.4 min. The calibration curve was linear up to 40g/mL. The mean recovery for benzhexol hydrochloride is 101.16. The assaywas precisewithin day and between days. The method provided excellent sensitivity, recovery, accuracy and reproducibility in therapeutic or toxic concentrations. Common excipients do not interfere.